Measurements and Patient Worksheets (Reports)
Note
11-44
Manual NT: Same as standard distance measurement: '2D Mode Measurements'
page 11-5 )
SonoNT: Switch to SonoNT by using the [NT Method] flipswitch and follow the
instruction:
1.
Press the [Calc] key on the control panel and choose [Early Gest.].
2.
Select the [NT] key.
3.
Use the trackball to move the cursor to the start point of the measurement and press the
right or left trackball key [Set] to fix the marker. A second cursor appears.
4.
Move the second cursor at the diagonal edge of the NT ROI to the second point of the
measurement and press [Set] again.
To re-adjust the start point, press the upper trackball key [Change] before completing the
measurement. This alternates the control from one cursor to the other.
5.
If the analysis finds a result: The NT measurement will be displayed. The NT position
can now be changed, using the [NT position] flipswitch. If the analysis can not find a
result: A warning appears on the screen: No valid NT-distanz found! Start over with step
4.)
on
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4