Download Print this page

GE Voluson S6 Basic User Manual page 551

Hide thumbs Also See for Voluson S6:

Advertisement

Note
13.1.3 Histogram
13.1.3.1 2D Histogram
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4
All values can be set back to default values when selecting a default monitor setting.
With this function the gray scale or color distribution within a marked Region of Interest (ROI)
will be graphically displayed. Three histograms may be shown on the screen simultaneously.
There are three possibilities to calculate the gray scale or color distribution:
(chapter '2D Histogram' on page 13-5 )
2D Histogram
(chapter '3D Histogram' on page 13-7 )
3D Histogram
(chapter 'Volume Histogram' on page 13-7 )
Volume Histogram
Operation:
1. Store a 2D mode, CFM mode or PD mode image.
2. Switch on Histogram by selecting [Utilities] and [Histogram].
The screen changes to the Histogram menu.
3. Select the number of histogram: 1, 2 or 3.
4. Use the trackball to place the rectangle over the ROI.
Utilities and System Setup
13-5