2.5.2.1 Electric Installation
2.5.2.2 Moving or lifting the System
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4
Do not operate the system in the vicinity of a heat source, of strong electric or magnetic fields
(close to a transformer), or near instruments generating high-frequency signals, such as HF
surgery. These can affect the ultrasound images adversely.
In the event the equipment has been brought from a cold environment (stock room, airfreight)
into a warm room, allow several hours for temperature balance and passing of condensation
humidity before switching on for the first time.
Do not cover the ventilation holes of the Voluson® S6/S8!
The system must be exclusively installed in medically used rooms. The equipment conforms
with regulations for electrical safety (IEC 60601) and safety class IIa according to the MDD
93/42/EWG regulation for use on humans patients. Probes are rated Type BF. Local safety
regulations may require an additional connection between the potential equilibrium bolt and
the building's grounding system.
Before switching on the first time, the local main voltage and frequencies have to be checked
against the values indicated on the Voluson® S6/S8 rating plate located on the rear panel.
Only authorized personnel must perform any change to the system.
The minimum required house installation must have 16A.
Do not detach power cord from Voluson® S6/S8. To avoid risk of electric shock, Voluson®
S6/S8 must only be connected to a supply mains with PROTECTIVE EARTH.
Do not connect any unauthorized equipment between power cord plug and wall power outlet.
Moving the system on planes
Moving the system on inclines *