Volume Mode
Note
9-18
Press the upper trackball key to change the function of the trackball from position to size or
vice versa.
6. Change the size of the Volume box by moving the trackball.
Moving:
decrease the box size in vertical direction
increase the box size in vertical direction
increase the box size in horizontal direction
decrease the box size in horizontal direction
low:
Fast speed / low scan density This mode is selected only
in case of expected movement artefacts. A loss of volume
resolution will result.
mid 2:
Standard VOL scan / medium scan density
max:
Slow speed/ high scan density
7. Set the volume sweep angle by using the right control below the menu area.
8. Select the Quality. This function changes the line density against acquisition speed.
9. To start 3D acquisition press the [Freeze] key, respectively the right trackball key (Start ->
displayed in the Status bar area on the monitor).
The volume acquisition starts, then the acquired images are displayed.
on page 9-20
'During 3D Acquisition'
If CRI is enabled in 2D Mode, it is also used in 3D/STIC pre mode and during 3D/STIC
acquisition. The settings (CRI value) are taken from the 2D settings. Use of CRI is indicated in
the info block. It is also possible to combine CRI with 3D/STIC Color.
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4