Safety
2.5.2 Environmental Conditions for Operation
Temperature:
Humidity:
Barometric pressure:
Light conditions:
Maximum operating altitude:
Pollution degree:
Overvoltage category:
Material group:
Total audible noise emission:
Note
2-8
Do not touch the patient and the Signal Input/Output lines (for example, USB port) at the same
time.
18˚C to 30˚C resp. 64˚F to 86˚F
30% to 80% RH, no condensation
700 to 1060 hPa
natural & artificial light source*
3000m; depending on the properties of the connected electronic
devices the maximum operating altitude is limited to the altitude stated
in the corresponding user manual of the connected electronic device
2
II
IIIB
<55dB
* Bright light could impact readability of screen.
Ultrasound systems are highly sensitive medical instruments that can easily be damaged by
improper handling. Use care when handling and protect from damage also when not in use.
DO NOT use a damaged or defective ultrasound system. Failure to follow these precautions
can result in serious injury and equipment damage.
This equipment is not to be used during transportation (e.g. ambulance cars, aircrafts).
Thermal Safety. Maintaining a safe thermal environment for the patient has been a design
priority at GE Healthcare. Software settings limit the power dissipated for the ultrasound
transducer as well as the motor-drive to values low enough to ensure that operating
temperatures stay below 43˚C.
Not to be used in sterile environment.
This equipment must not be used in oxygen enriched atmosphere or in the presence of
inflammable gases (e.g. anesthetic gases).
The use of the system outside the described conditions or intended use, and disregarding
safety related information is considered as abnormal use. The manufacturer is not liable for
damage caused by abnormal use of the device!
Use for diagnostic purposes only.
Voluson® S6/S8 Basic User Manual