Measurements and Patient Worksheets (Reports)
11.2.5.11 AV/LA (Aortic Valve/Left Atrium)
Note
11.2.5.12 PV (Pulmonary Valve)
Note
11.2.5.13 Cardiology Calculations in CFM Mode
11.2.5.14 PISA
11-68
To measure the Aortic Diameter or the Left Atrium Diameter in 2D mode:
1.
Press the [Calc] key on the control panel.
2.
Select the item [AV/LA] and then the desired measurement parameter [Ao Root Diam] or
[LA Diam].
3.
Use the trackball to move the cursor to the start point of the measurement and press the
right or left trackball key [Set] to fix the marker. A second cursor appears.
4.
Move the second cursor to the second point of the measurement and press [Set] again.
To re-adjust the start point, press the upper trackball key [Change] before completing the
measurement. This alternates the control from one cursor to the other.
To measure the Diameter of the PV
1.
Press the [Calc] key on the control panel.
2.
Select the item [PV] and then the measurement parameter [PV Diam].
3.
Use the trackball to move the cursor to the start point of the measurement and press the
right or left trackball key [Set] to fix the marker. A second cursor appears.
4.
Move the second cursor to the second point of the measurement and press [Set] again.
To re-adjust the start point, press the upper trackball key [Change] before completing the
measurement. This alternates the control from one cursor to the other.
(chapter 'PISA' on page 11-68 )
PISA
To measure the PISA Radius of the Mitral Valve (MV), Tricuspid Valve (TV), Aortic Valve (AV)
and Pulmonary Valve (PV) in Color Flow Mode:
(Pulmonary Valve) in 2D mode:
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4