The selected Thermal Index is displayed in the [TI select] key.
Remarks:
•
While you are scanning, notice the index numbers you are using and which controls
affect the readings.
•
Try to keep the index numbers as low as you can, while maintaining diagnostic
information within the image. This is particularly important when scanning the fetus.
•
The biopsy lines must be programmed once by the service personnel or the user. The
procedure must be repeated if probes and/or biopsy guides are exchanged!
•
Before performing a biopsy make sure that the displayed biopsy line coincides with the
needle track (check in a bowl filled with approx. 47˚C warm water). For further
instructions review: To program a Single Angle Biopsy Line
Single Angle Biopsy Line'
'To program a Multi Angle Biopsy Line'
•
Please read the "Instructions for safe Use" in Probes and Biopsy / Biopsy Special
(chapter 'Biopsy Safety and Maintenance' on page 2-19 ) .
Concerns
Operation:
Select the [Utilities] and the [Biopsy Line] key. The Biopsy guideline appears on the monitor.
The Biopsy line is switched on if the [Biopsy Line] key is highlighted. To switch OFF the Biopsy
guideline select the [Biopsy Line] key again.
The [Biopsy Line] key toggles between ON and OFF.
•
To program a Biopsy guideline review: To program a Single Angle Biopsy Line
'To program a Single Angle Biopsy Line'
Line
(chapter 'To program a Multi Angle Biopsy Line' on page 5-16 )
Remarks:
•
For Handling, Sterilization, Mounting a biopsy guide, etc. review: Biopsy Special
(chapter 'Biopsy Safety and Maintenance' on page 2-19 ) .
Concerns
•
For each probe one biopsy guideline is programmable.
on page 5-15 ) To program a Multi Angle Biopsy Line (chapter
on page 5-16 )
on page 5-15 ) To program a Multi Angle Biopsy
(chapter 'To program a
(chapter