Download Print this page

GE Voluson S6 Basic User Manual page 417

Hide thumbs Also See for Voluson S6:

Advertisement

Note
Note
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4
1.
Select [Early Gestation] and then [GS] on the menu area. A cursor appears.
2.
Select "Triple Caliper" by turning the left rotary control.
3.
Use the trackball to move the cursor to the start point of the measurement and press the
right or left trackball key [Set] to fix the marker. A second cursor appears.
4.
Move the second cursor to the second point of the measurement and press [Set] again.
To re-adjust the start point, press the upper trackball key [Change] before completing the
measurement. This alternates the control from one cursor to the other.
5.
Measure the second distance as described above.
6.
Perform the measurement of the third distance in the same manner.
Method 2:
The result is displayed immediately after the measurement of the 1 Distance.
1.
Select [Early Gestation] and then [GS] on the menu area. A cursor appears.
2.
Select "Distance" by turning the left rotary control.
3.
Use the trackball to move the cursor to the start point of the measurement and press the
right or left trackball key [Set] to fix the marker. A second cursor appears.
4.
Move the second cursor to the second point of the measurement and press [Set] again.
The distance between the two points is displayed.
To re-adjust the start point, press the upper trackball key [Change] before completing the
measurement. This alternates the control from one cursor to the other.
Measurements and Patient Worksheets (Reports)
11-41

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Voluson s8