Safety
2.6.7.5 Planned Maintenance
2.6.7.6 Environmental requirements for Probes
2-18
The following maintenance schedule is suggested for the system, probe and reusable biopsy /
biopsy bracket to ensure optimum operation and safety.
Do the Following
Inspect the Probes
Clean the Probes
Disinfect Probes
After cleaning and disinfection, inspect the probe's lens, cable, casing and connector. Look for
any damage that would allow liquid to enter the probe. Also, inspect the probe functionality by
live scan. If any damage is found, do not use the probe until it has been inspected and
repaired/replaced by a GE service representative.
Probes must be operated, stored, or transported within the parameters outlined below.
Ensure that the probe face temperature does not exceed the normal operation temperature
range.
Avoid temperatures above 50ºC
Daily
After Each Use
X
X
X
As Necessary
X
X
X
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4