Philips HEARTSTART XLT Service Manual page 102

Advertisement

External Assemblies
4-16
Replacement
1. Ensure the lower edge of the battery PCA is seated in its slot in the case
and that the battery spring is in its slot. Correct orientation of the Battery
Eject Spring is angled slightly toward the Battery PCA, and NOT angled
up out of the battery compartment. Adjust the angle of the spring as
needed by turning it while keeping it in place in its slot.
2. Carefully insert a battery in the battery compartment. Be sure the battery
stays flat and does not push up out of position.
3. While keeping the battery in place, replace the battery cover starting with
the end by the battery PCA. See Figure 4-9, which shows the case stand-
ing upright for clarity. Insert the tab on the cover into the mating slot next
to the battery spring. Then lower the cover into position, allowing the
square post on the cover to mate with the recess in the case next to the
Power Connector.
4. Replace all screws and tighten.
Figure 4-9 Installing the Battery Cover
Battery PCA
Slot in cover
Tab on cover
Removal and Replacement