Momentary Messages
Table 3-6 Extended Self Test Failures
Failure
RAM
ROM
System
CODEC
IRDA
Timebase
System
FAIL 10
Data Card
Pacer
Defib
FE
SpO2
3-18
Extended Self Test Failures
Should the unit report a FAIL in the Extended Self Test, resolve it using the
solutions below.
Possible Cause
Failure on Control PCA.
Lithium backup battery failure.
Data Card full, or incompatible, or
defective.
Failure on Control PCA.
Failure on Power PCA.
Failure on Control PCA.
Failure on Control PCA.
Failure on Power PCA.
Failure on Interface PCA.
Failure on SpO 2 PCA.
Failure on Control PCA.
Suggested Solution
Replace Control PCA.
Replace lithium battery.
Replace Data Card with new, empty
M3510A/M5510A.
Replace Control PCA.
Replace Power PCA.
Replace Control PCA.
Replace Control PCA.
Replace Power PCA.
Replace Interface PCA.
Replace SpO 2 PCA.
Replace Control PCA.
Troubleshooting