Table 3-3 System Messages
Message
Attach Pads Cable
The pads cable is not properly attached to
the device.
Configuration Lost
The configuration is reset to the default
settings.
Data Card Disabled
The PC card is not in use because it is full,
incompatible, absent, or inserted after the
M3500B/M5500B was turned on.
ECG Fault
The ECG data acquisition system failed
and data is unavailable from the 3- or 5-
lead monitoring electrodes.
Monitor Failure - Cycle
An error has occurred in the monitoring
subsystem.
Power
Low Battery
The battery has sufficient capacity remain-
ing to provide only about ten minutes of
monitoring time and six shocks before the
M3500B/M5500B shuts off.
Leads Off
The monitoring electrodes are not applied
or are not making proper contact with the
patient.
The ECG cable is not connected.
The internal cable from the ECG connec-
tor to the Interface PCA is disconnected
Failure on the Interface PCA.
No Pads
The multifunction defibrillation electrode
pads are not properly connected to the
M3500B/M5500B.
M3500B/M5500B Service Manual
System Messages
System messages remain on the display until the specified action is taken or
no longer relevant. They are intended for the end user and appear in the
Instructions for Use. They are duplicated here for the reference of the service
person.
Before replacing any parts, check to see if the cables and flex circuits are
properly connected. See "Servicing Notes" on page 4-1.
Description
System Messages
Corrective Action
Check the cable connection
1. Reconfigure the M3500B/M5500B.
2. If problem reoccurs, service the unit.
See "Troubleshooting" on page 3-1 and
Table 3-6 on page 3-18.
If possible, turn the M3500B/M5500B off
for more than 2 minutes, then insert a
new, empty, M3510A/M5510A Data Card
and turn the device on.
Service the unit. See "Troubleshooting" on
page 3-1 and Table 3-7 on page 3-19.
Turn power off, then on. If message reap-
pears, service the unit.
See "Trouble-
shooting" on page 3-1.
Replace the battery with a fully charged
M3516A/M5516A battery.
Check that the monitoring electrodes are
properly applied.
Check that the ECG cable is properly con-
nected.
Check the internal cable.
Replace the Interface PCA.
Check the pads cable connection.
3
3-13