Top Case Assemblies
ECG Connector
SpO 2 Connector
Interface PCA Connectors:
SpO 2
ECG
4-38
SpO
PCA
2
The SpO
PCA is located next to the ECG and SpO
2
Removal
1. Disconnect and remove the ECG connector. See "ECG Connector" on
page 4-40.
2. Disconnect and remove the SpO
page 4-42.
3. Note the orientation and placement of the flex circuit from the SpO
to the Interface PCA. Disconnect this flex circuit at both ends. These con-
nectors have a sliding bar latch - slide out toward the edge of the PCA to
release, then pull out the flex.
4. Remove the SpO
PCA and discard the flex circuit to the Interface PCA.
2
Figure 4-22 Removing the Screws - SpO
Interface PCA
connector. See "SpO
2
PCA
2
Ferrite
Flex Circuit - SpO 2
SpO 2
PCA to Interface
PCA
PCA
connectors.
2
Connector" on
2
PCA
2
Removal and Replacement