Bottom Case Assembly
NOTE
4-66
Bottom Case Assembly
The bottom case assembly consists of the plastic case, with the Data Card
Door, Battery Cover, and Battery Eject Latch mechanism all pre installed at
the factory. The bottom case must be replaced only when there is damage to
the case itself; everything else in the bottom case is individually replaceable.
Parts to be moved from old bottom case to new
The following subassemblies must be removed from the old top case and
installed into the new one. Follow the Removal and Replacement instructions
in this chapter for each subassembly. Remove them in the order shown, and
replace them in reverse order.
1. Battery PCA, spring, cable (see"Battery PCA" on page 4-58).
2. Battery eject spring (see "Battery PCA" on page 4-58).
3. Patient Connector, cables, hardware (see "Patient Connector" on page 4-
65).
4. Power Connector, cables, hardware (see "Power Connector" on page 4-
64).
5. Defibrillator capacitor (see "Defibrillator Capacitor" on page 4-60).
6. Power PCA and shield (see "Power PCA" on page 4-62).
7. Battery.
8. All screws.
New parts to be installed - must be ordered separately
In addition, the following parts cannot be reused from the old case, and must
be ordered separately and installed into the new bottom case:
Case Label Set
These are available in all the supported languages. See the
"Replacement Parts" chapter,5Table 5-12 to select the correct
language.
Primary Label
When replacing the bottom case, the Primary Label containing the
unit's serial number must also be replaced. Be prepared to provide
information identifying the unit when ordering the bottom case.
To enable device tracking as mandated by US Federal law, the Primary Label
must be applied before placing the unit back into service.