Table 2-1 Performance Verification and Safety Tests
Test or Inspection to
Test Group Name
Inspect unit, accessories,
Visual Inspection
cables, etc. as described on
(VI)
page 2-12.
In normal Operating Mode,
Functional
perform the following func-
Checks (F)
tional checks:
• ECG (page 2-13).
• Shock Advisory (page 2-
14).
• Synchronized Cardiover-
sion (page 2-14).
• SpO 2 (page 2-15).
In Diagnostic Mode, run the
Extended Self
Extended Self Test (page 2-
Test (X)
17). Includes Data Card Test
and time/date check.
M3500B/M5500B Service Manual
Test and Inspection Matrix
The matrix in Table 2-1 summarizes performance verification tests and
inspections for the M3500B/M5500B; including test name, test or inspection
to perform, expected test results, and data to record.
Expected Test Results
Perform
• If no unusual damage, no
corrosion, no missing
items,
then Visual Inspection
passes
• If all functions respond as
expected:
• Waveform clear on dis-
play; HR correct on display;
HR alarm works. Leads off,
pads off indicators as
expected.
• Shock Advised only when
appropriate
• Shock delivered with cor-
rect timing (< 60msec)
• 95% -100%
then Functional test passes
If "Pass" reported on all
tests applicable to the device
configuration and options,
then Extended Self test
passes
Test and Inspection Matrix
Data to Record
P (pass) or F (fail)
Example VI:p
Example F:p
Example X:p