2 Pin Connector to
Battery PCA
4 Pin Connector to
Battery PCA
High Voltage Spade
Connectors (2)
9 Pin Connector
Patient Connector
Power Connector
(hidden)
M3500B/M5500B Service Manual
Figure 4-33 Disconnecting the Power PCA
Replacement
1. Place the new Power PCA into position in the case. Replace the 6 T10
screws and tighten.
1. Replace and reconnect the defibrillator capacitor. See "Defibrillator
Capacitor" on page 4-60.
2. Reconnect the 3 connectors from the Patient Connector (2 individual
High Voltage, 1 multipin).
3. Reconnect the 2 connectors from the Battery PCA.
4. Reconnect the 1 connector from the Power Connector.
5. Reconnect the large ribbon cable to the Control PCA. Replace and relatch
the metal bail.
Bottom Case Assemblies
Large Ribbon Cable
4-63