Philips HEARTSTART XLT Service Manual page 202

Advertisement

Specifications
7-6
Noninvasive Pacing
Waveform: Monophasic Truncated Exponential.
Current Pulse Amplitude: 10 mA to 200 mA (5 mA increments); accuracy
+
10 mA - 50 mA
5 mA, 50 mA - 200 mA
Pulse Width: 20 ms with accuracy +0, -5 ms.
Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy
Modes: Demand or Fixed Rate.
Refractory Period: 340 msec (30 to 80 ppm); 240 msec (90 to 180 ppm).
SpO
/Pulse Oximetry
2
Accuracy with:
M1191A sensor - 1 standard deviation 70% to 100%,
M1192A sensor - 1 standard deviation 70% to 100%,
M1194A sensor - 1 standard deviation 70% to 100%,
NELLCOR sensors - 1 standard deviation 80% to 100%
Resolution: 1%.
SpO
Alarm Limits: Three preset low alarm limits: 90, 85, and 80%, default
2
is off.
INOP Alerts: Triggered by disconnected sensor, noisy signal, light interfer-
ence or low signal (non-pulsatile).
Event Storage
Internal Event Summary:
The internal Event Summary stores up to 300 events and up to 50 waveforms.
Events can be marked with a Mark Event symbol and, if configured for drug
annotation, the following labels can be added: Epinephrine, Atropine,
Lidocaine, or Other.
The Print Summary key on the front panel is used to print the internal Event
Summary.
Data Card Event Summary:
The Data Card stores continuous ECG waveforms and events on a Type II
PCMCIA card, SanDisk SDP3B 8MB ATA FlashDisk. One 8MB card can
store approximately 2 hours of continuous ECG waveforms and events.
+
10%.
+
1.5%.
+
2.5%.
+
2.5%.
+
4.0%.
+
3.0%.
Specifications