External Assemblies
Platen
Platen Springs
Hinge Pins
4-12
Figure 4-7 Removing the Printer
Sliding
Door
Replacement
1. The replacement printer ships with a grey sliding door and a grey platen
installed. They must be replaced with the yellow sliding door and yellow
platen that came in the printer kit. Refer to "Printer Sliding Door" and
"Printer Platen" on page 4-13.
2. Open the paper door on the new printer. Leave the paper door open until
the screws under the label are tightened.
3. Put the 4 screws in the holes in the new printer. Get the 2 screws back in
the holes under the label with tweezers or fine nose pliers.
4. Reconnect the ribbon cable. Be sure to engage the 2 latches on the ends of
the ribbon connector.
5. Lower the printer straight down into the instrument.
6. Tighten all screws.
7. Clean the surface where label will be placed with alcohol. Allow it to dry.
8. Replace the label.
Printer
ribbon
cable
Removal and Replacement