M3500B/M5500B Service Manual
New parts to be installed - provided with top case
The Replacement Top Case Assembly comes with the following new compo-
nents, which must be installed in the field. Do not reuse the old parts.:
ECG/SpO
connector gaskets (not available separately)
2
ECG/SpO
connector retaining clips (M2475-07101)
2
New parts to be installed - must be ordered separately
In addition, the following parts cannot be reused from the old case, and must
be ordered separately and installed into the new top case:
Branding Label Set (includes the Ambient Light sensor cover)
Philips label set (M3500-69556)
a.
Laerdal North America label set (M3500-69557)
b.
Laerdal Other label set (M3500-69558)
c.
Manual Keypad
These are available in 2 versions (pacing, no pacing) and in all
the supported languages. See the "Replacement Parts"
chapter,5Table 5-7 and Table 5-8 to select the correct version and
language.
Case Label Set
These are available in all the supported languages. See the
"Replacement Parts" chapter,5Table 5-12 to select the correct
language.
Top Case Assembly
4-57