Download Print this page

Philips HEARTSTART XLT Service Manual page 135

Advertisement

M3500B/M5500B Service Manual
Peel the backing material off the rubber overlay. Handle carefully
b.
and touch the adhesive as little as possible.
Starting with the top edge, align the overlay exactly and roll it down
c.
slowly into place.
Press firmly all over, especially the edges, to adhere the new overlay
d.
into place.
8. Replace the Manual Door.
9. Be sure to specifically check the operation of each of the keys during Per-
formance Verification by performing a Controls Test (see "To check the
Controls (keys):" on page 2-19).
Top Case Assemblies
4-49
4

Advertisement

loading

This manual is also suitable for:

M3500bM5500b