Philips HEARTSTART XLT Service Manual page 213

Advertisement

WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
CAUTION
M3500B/M5500B Service Manual
Do not immerse, or pour fluids on, any portion of the M3500B/M5500B.
Do not use the M3500B/M5500B in a flammable or oxygen-rich atmosphere. This
can cause an explosion hazard.
Avoid connecting the patient to several devices at once. Leakage current limits may
be exceeded. Do not use a second defibrillator on the patient while pacing with the
M3500B/M5500B.
Avoid contact between the patient and metal objects, such as the gurney. Contact
with metal objects could cause unintentional current pathways.
Operating the M3500B/M5500B or its accessories in conditions outside the environ-
mental specifications can result in device or accessory malfunction.
Avoid exposing the display to direct sunlight on a hot day. Overheating can occur
and cause the display to black out, making the M3500B/M5500B temporarily unus-
able.
Medical electrical equipment which does not incorporate defibrillator protection
should be disconnected during defibrillation.
Electric shock hazards exist internally. Do not remove assembly screws. Refer ser-
vicing to qualified personnel.
This device has not been evaluated for use with electrosurgery equipment.
Safety Considerations
7-17