M3500B/M5500B Service Manual
Sp0
Functional Check
2
This check only needs to be performed if SpO
1. Attach the SpO
transducer to your finger.
2
2. Activate Manual Mode and press the SpO
3. The SpO
value displayed should be in the range of 95-100%. If the value
2
is less than 95%, check that your finger is fully inserted into the sensor
and properly positioned.
Performance Verification