Download Print this page

Philips HEARTSTART XLT Service Manual page 203

Advertisement

M3500B/M5500B Service Manual
General
Dimensions: 4.05" (H) x 11.25" (W) x 13.65" (D).
103 mm x 286 mm x 347 mm.
Weight: Standard Configuration weighs < 10 lbs (4.5 kg) including battery,
full roll of paper, defibrillator patient cable, 1 set of pads, and without the car-
rying case; with the carrying case, it weights < 14 lbs (6.4 kg).
Environmental
o
o
Temperature: 0
to 50
C operating, -20
Thermal paper may darken above 55
Charging the battery at temperatures above 35
life.
Storing the battery for extended periods at temperatures above 40
will reduce battery capacity and degrade battery life.
Operating and storage specifications for electrodes may vary. Refer to
the manufacturer's specifications for details.
Humidity:
Up to 95% Relative Humidity
Printer paper may jam if paper is wet.
Thermal printer may be damaged if wet paper is allowed to dry while
in contact with printer elements.
Altitude:
Operating: up to 15,000ft.
Shipping: up to 20,000 ft.
Shock (drops onto concrete): Unit survives 39" (1 m) drops onto all surfaces
(faces, corners, and edges) enclosed in carrying case. Exposed handle surfaces
survive 24" drops.
Vibration: Mil Std. 810E 514.4 Category 6 Helicopter, General Storage,
UH60.
Water Resistance: IPX4, splash proof per IEC 60529, with Data Card door
closed.
Safety: Meets IEC 60601-1 (EN 60601-1), UL 2601-1,
CAN/CSA C22.2 NO.601.1
EMC: Meets EN 60601-1-2.
Other Considerations: Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation: Continuous.
Specifications
o
o
to 70
C storage.
o
C.
o
C may degrade battery
o
C
7
7-7

Advertisement

loading

This manual is also suitable for:

M3500bM5500b