Top Case Assemblies
Power PCA
Cable
Interface
PCA Cable
Metal
Bails
Speaker
Cable
Display Backlight
Cable
4-32
To the Rest of the Unit
1. Connect the speaker cable to the Control PCA.
2. Connect the printer cable.
3. Connect the 2 large ribbon cables - 1 from the Power PCA, 1 from the
Interface PCA. Be sure to replace the 2 metal bails on the 2 large ribbon
connectors. If the clips have become loose or are missing, replace them.
Figure 4-18 Control PCA Connections