M3500B/M5500B Service Manual
SYMBOLS: Mark Triangle (for presses of the Mark key), an Alarm Bell
(Alarm Limit Violations), Lightening Bolt (Shock Delivered), Vertical stripe
Boundaries/Pacer/Sync Tick Marks).
Event Printing: Mark Event key automatically documents events and ECG
during defibrillation episodes. The Mark Event key can annotate the event
with one of the following labels: Epinephrine, Atropine, Lidocaine, and
Other.
Auto Printing: The printer can be configured to automatically print on Mark,
Charge, Shock and Alarm.
Delayed Printing: The printer can be configured to run real time or with a six
second delay.
Reports: The following can be printed: Event Summary, Configuration,
Extended Self Test, System Log.
Speed: 25 mm/s with an accuracy of
+
Amplitude Accuracy:
Paper Size: 50 mm by 30 m (100 ft.).
Battery and Battery Power Modules
Battery Type: 2 Ah, 12V, rechargeable, Sealed Lead Acid.
Battery Dimensions: 2.4" (H) x 0.94" (W) x 7.2" (D).
61.7 mm (H) x 23.9 mm (W) x 182 mm (D).
Weight: 1.4 lb. (0.65 kg).
Charge Time:
Approximately 14.5 hours to 100%, indicated by LED on power
module.
Approximately 2.5 hours to 90%, indicated by LED on power mod-
ule.
Capacity: 2.7 hours ECG monitoring or 50 full-energy discharges or 2.1
hours ECG monitoring while pacing (with a new, fully charged battery and
o
temperature above 20
C).
Battery Indicators: LOW BATTERY message appears on monitor when at
least 10 minutes of monitoring time and 6 maximum-energy discharges
remain (with a new battery and temperature above 20
Battery Storage: Should not be stored above 40
time.
Charger Output: Unit can be operated using Class I AC or DC power mod-
ule alone with no battery installed.
+
5%.
+
10% or
50 uV.
o
C for extended periods of
Specifications
o
C).
7-5
7