User-replaceable Parts and Accessories
4-4
User-replaceable Parts and Accessories
The Instructions for Use gives detailed instructions on replacing the follow-
ing:
Patient cables & sensors
Battery
Printer paper
For the convenience of the service person, instructions on replacing the Man-
ual Door and the Carrying Case are duplicated here.
Manual Door
The following instructions detail how to remove and replace the Manual
Door. Refer to Table 5-6 on page 5-9 to select the correct replacement.
Removal
1. Open the door to a vertical position as shown in Figure 4-1.
2. Pull the hinge end of the door away from the case. It will snap free.
Replacement
1. If you are installing a new door, install the new labels as follows
Clean both top and bottom door surfaces with isopropyl alcohol.
a.
Allow them to dry.
Lay the door on a clean flat surface with its underside facing up (the
b.
side that is concealed when the door is closed).
The new door will come with labels both for units with Pacing and
c.
without Pacing. Select the correct label for your unit.
Peel the new Manual Operation label off the backing material.
d.
Starting with the top edge, align the label carefully with the recess on
e.
the door and roll it down slowly into place. When installed, it should
be as shown in Figure 4-3.
Press firmly all over, especially the edges, to adhere the new label
f.
into place.
Turn the door over to expose its top surface.
g.
Peel the new Manual label off the backing material.
h.
Starting with the top edge, align the label very carefully with the
i.
recess on the door and roll it down slowly into place.
Press firmly all over, especially the edges, to adhere the new label
j.
into place.
2. Align the door with the door hinges as shown in Figure 4-1.
Removal and Replacement