Diagnostic Tests
2-26
7. Using the
Rate
8. Using the
Output
9. The Pacer should be delivering a current of 200mA +20mA. Record the
delivered current indicated by the Pacer tester as "bbb".
10. The display on the M3500B/M5500B should read the delivered current as
measured by the Pacer tester ("bb") +20 mA.
11. Turn off Pacing by pressing the
button, increase the rate to 180 bpm.
button, increase the output to 200 mA.
button.
Pacer
Performance Verification and Safety Tests