M3500B/M5500B Service Manual
Error Codes
The System Log provides error codes in 5 digit hexadecimal format. Before
replacing any parts, check to see if the cables and flex circuits are properly
connected. See "Servicing Notes" on page 4-1.
Confirm any errors by running the self-tests again, making sure that no one
touches the unit during the test.
Meaning
Defib failure - charging circuits.
Defib failure - biphasic processor.
Leads front end failure.
Pacer failure.
Processor error.
System monitor failure.
Processor Synchronization Error
Synchronization Time Out
Front end failure.
Pacer failure.
Monitor processor failure.
SpO 2 problem.
Advisory Failure.
Error Codes
Possible Solutions
1. Replace Power PCA.
2. Replace Control PCA.
1. Replace Power PCA.
2. Replace Control PCA.
1. Replace Interface PCA.
2. Replace Control PCA.
Replace Power PCA.
Replace Control PCA.
1. Cycle power.
2. Replace Control PCA.
Cycle power.
Cycle power.
• If error does not repeat, run ALL
Performance Verification Tests. If
unit passes all tests, the unit can
be returned to service.
• If error repeats, replace Control
PCA.
1. Replace Control PCA.
2. Replace Interface PCA.
1. Replace Power PCA.
2. Replace Control PCA.
1. Replace Control PCA.
2. Replace Interface PCA.
1. Replace SpO2 PCA.
2. Replace Interface PCA.
3. Replace Control PCA.
Replace Control PCA.