M3500B/M5500B Service Manual
Replacement
1. The new capacitor will come with a shorting bar connecting the two ter-
minals. Disconnect the shorting bar.
2. Slide the wires through the bottom piece of foam, routing them like the
original wires.
3. Plug the terminals onto the spade connectors on the Power PCA. Note the
correct polarity is with the red wire near the red inductor and the white
wire near the back corner of the Power PCA.
4. Assemble the top and bottom foam pieces around the capacitor and slide
the assembly down in place on the Power PCA.
Bottom Case Assemblies
4-61