Table 2-1 Performance Verification and Safety Tests (Continued)
Test or Inspection to
Test Group Name
Using only the AC or DC
Defibrillator Test -
power module, enter Diag-
AC or DC Power
nostic Mode and run the
Module (DM)
(Required only
Defibrillator Test (at
when AC or DC
200Joules) (page 2-27):
power module
used.)
Displayed by
M3500B/M5500B:
• Available Energy after
Shock
• Msec to charge
• Delivered energy
• Impedance
• Defib errors
Using only battery power,
Defibrillator Test -
enter Diagnostic Mode and
Battery Power
run the Defibrillator Test (at
(DB)
200Joules) (page 2-28).
Displayed by
M3500B/M5500B
• Available Energy after
Shock
• Msec to charge
• Delivered energy
• Impedance
• Defib errors
Enter Diagnostic Mode and
Defibrillator
run the Defibrillator Disarm
Disarm Test (D)
Test (page 2-29)
M3500B/M5500B Service Manual
Expected Test Results
Perform
If the measurements are as
follows:
• 0
• < 15000 msec
• Actual delivered energy
200 +7%
• 42 to 57 Ω
• None (0)
then the Defibrillator test
(Power Module) passes
If the measurements are as
follows:
• 0
• < 3000 msec
• Actual delivered energy
200 +7%
• 42 to 57 Ω
• None (0)
then the Defibrillator test
(Battery Power) passes
If all readings as expected
Available energy - failure >0
No errors reported
then the Defibrillator Disarm
test passes
Test and Inspection Matrix
Data to Record
P (pass) or F (fail)
Example: DM:p
Example: DB:p
Example: D:p
2-5
2