Philips HEARTSTART XLT Service Manual page 128

Advertisement

Top Case Assemblies
CAUTION
4-42
SpO
Connector
2
The SpO
Connector is located in the corner of the top case, next to the ECG
2
Connector.
See cautions regarding handling and connection of flex circuits on page 4-1.
Removal
1. Disconnect and remove the ECG Connector. See "ECG Connector" on
page 4-40.
2. Remove the 2 screws from the SpO
3. Lift up the PCA gently. Pivot to make the underside visible as shown in
Figure 4-23.
4. Note the orientation and placement of the flex circuit between the SpO
connector and the SpO
no latch - just pull it straight out.
5. Leave the flex circuit from the SpO
nected.
6. Release the locking tabs on the metal clip around the SpO
using a small flat-bladed screwdriver. Pull up and remove the metal clip.
See Figure 4-25.
7. Slide the connector, gasket, and flex circuit out through the hole in the
case.
Figure 4-25 Removing the SpO 2 Connector
SpO
Connector
2
PCA. See Figure 4-22.
2
PCA. Disconnect it at the SpO
2
PCA to the Interface PCA con-
2
Metal Clip
2
PCA. There is
2
connector
2
SpO2 PCA
Removal and Replacement