Download Print this page

Philips HEARTSTART XLT Service Manual page 59

Advertisement

Firmware Versions
Main:
DSP:
196:
Key:
SpO2:
Language:
Serial Number:
Options:
Shocks:
M3500B/M5500B Service Manual
2. Select Print Log and press
3. The printer will print the System Log strip. See Figure 3-2.
4. Check the printout to verify the printed results are consistent with the
hardware in place. Verify options installed (SpO2, pacing) and the unit's
serial number (on the bottom of the case). If the printout is not correct,
investigate and resolve the source of the mismatch. Then reset the hard-
ware options and serial number as needed using the Support Software
Tool. See "The Language Support Tool" on page 4-67.
5. If there are errors reported in the System Log:
Check the time and date stamps to see if they are recent errors or not.
a.
Consult Table 3-2 "Error Codes" on page 3-11 to identify the errors.
Then begin troubleshooting as needed (see the "Troubleshooting"
chapter.)
Check that the time and date are correct. If they are not correct, access
b.
Configuration Mode and reset them. See "Accessing Configuration
Mode" on page 3-5.
Figure 3-2 Sample System Log Printout
SYSTEM LOG
Error Codes
34
90007 18:33
02
57.00
06
02.42 01.04
English
US0100241
Pacer
SpO2
2
Diagnosing Internal Failures
.
ENTER
19 Jan 2000
3
3-7

Advertisement

loading

This manual is also suitable for:

M3500bM5500b