Philips HEARTSTART XLT Service Manual page 21

Advertisement

2
Performance Verification and Safety Tests
Overview
This chapter describes the tests and inspections required to verify perfor-
mance of the M3500B/M5500B Defibrillator/Monitor.
The information is presented in the following sequence:
Test Matrix
Test Equipment
Performance Verification
Mandatory Testing
The Performance Verification Tests in this chapter are intended to verify
proper operation of the M3500B/M5500B following repair. The level of test-
ing required corresponds to the type of repair performed, and is divided into 3
categories:
External/No Trouble Found
Printer
Internal
External Repairs/No Trouble Found
External Repairs are those involving the repair or replacement of one or
more of the items below. No Trouble Found applies when no malfunction
can be found, or when the problem appears to be due to improper use. In
either situation, the key point is that the case has not been opened.
ECG cable
SpO
cable or sensor
2
Battery
A list of the tests required to verify per-
formance of the M3500B/M5500B,
along with expected test results.
A list of commercially available test, cal-
ibration and simulation devices needed
to perform performance verification
tests.
Step-by-step instructions for performing
each of the Performance Verification
Tests.
2
2-1