Mandatory Testing
2-2
Labels
AC or DC Power module
Consumables (ECG monitoring electrodes, multifunction defibrilla-
tion pads, printer paper)
Main fuse (on Battery PCA)
Manual Door Assembly
Data Card Door Assembly
Battery Eject Assembly
The following testing is required after an External Repair or when the out-
come of the service is No Trouble Found (when the case has not been
opened):
Perform the Visual Inspection (page 2-12).
Run the Extended Self Test (page 2-17).
Print and Verify the System Log (page 2-16).
Printer
If the printer was replaced, and the case was not opened, the following tests
are required:
Perform the Visual Inspection (page 2-12).
Run the Extended Self Test (page 2-17).
Run the Printer Test (page 2-21).
Print and Verify the System Log (page 2-16).
Internal Repairs
If the case was opened (regardless of what the repair involved), all of the Per-
formance Verification Tests must be performed, beginning with "Visual
Inspection" on page 2-12.
Performance Verification and Safety Tests