System Functional Descriptions
Indicator Functions
All panel LEDs (Manual, Sync, Pacer) are controlled by the Control PCA.
The LEDs are connected into the Interface PCA via flex circuits; they are
then routed to the Control PCA via a ribbon cable.
6