Specifications
7-4
ECG Size: 2.5, 5, 10, 20, 40 mm/mV.
Frequency Response:
AC Line Filter: 60 Hz/50Hz.
Pads ECG for Display: Monitor (.15-40 Hz)/EMS (1-30 Hz).
Pads ECG for Printer: Monitor (.15-40 Hz)/EMS 1-30 Hz).
Leads ECG for Display: Monitor (.15-40 Hz)/EMS (1-30 Hz).
Leads ECG for Printer: Diag (.05-150Hz)/EMS (1-30)/Monitor (.15-40
Hz).
Patient Isolation:
ECG: Type CF
SpO
: Type CF
2
Defib: Type BF (multi-function defibrillator electrode pads)
Display
Display Size: 96 mm x 72 mm.
Display Type: Transflective active matrix LCD with LED backlight.
Display Resolution: 320 x 240 pixels.
Sweep Speed: 25 mm/s nominal (stationary trace; sweeping erase bar).
Viewing Time: 4 seconds.
Thermal Array Printer
Continuous Real Time Strip:
User starts and stops the print strip. The Print Strip prints the selected ECG
lead with the following data:
HEADER 1: Date, Time, Heart Rate, the SpO
text "Delayed" if the recording has been configured for Delayed Mode. Prints
every 12 seconds.
HEADER 2: Current mode (AED/Manual) the current Lead, the current
Gain, filter setting, the text "Sync" if Sync has been enabled, and the Pacer
Settings (consisting of the Pacer Mode, Rate, and Current (if currently pacing
the patient). Prints every 12 seconds, with Header 1.
HEADER 3: Changes in Mode, Gain, Lead, Sync, and Pacer Settings.
FOOTER: Drug Annotations, HR/SpO
of an Analysis in AED Mode (No Shock Advised, Shock Advised, or Cannot
Analyze), Charging to xxxJ, Shock Delivered, No Shock Delivered, Disarm,
Battery Low, Battery Critical.
Value (if available), and the
2
limits on a Limit Alarm, the Results
2
Specifications