Customer Supplied Displays - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Table 2-6. Electrical Specifications of Database Server Components
Philips Component
UPS for Database Server
(1000 VA, 230V)
Power Distribution Module
Conversion factors for power dissipation are the following:
Conversion factor for air conditioning requirements for power dissipation is:
Customer
The following table gives recommended requirements for displays and video cables connected
Supplied
to Information Centers and Clients.
Displays
Note
For displays and video cables not supplied by Philips as part of an Information Center system:
Table 2-7. Recommended Display Specifications for Information Center Displays
Specification
Screen Resolution
Vertical Refresh Rate
Product
Part #
M3883-60001
M3166-60000
BTU/hr = Watts x 3.413
Kcal/hr
= Watts x 0.8598
air conditioning
1 ton = 8900 BTU/hr = 3024 Kcal/hr
The support user is responsible for all aspects of their selection, purchase, installation,
repair, and disposal
Philips cannot assure their system compatibility
Philips cannot assure compliance with the ECG aspect ratio and 25 mm/s +10% sweep
speed specification of Information Center software, which complies with the ANSI/
AAMI EC-13 Standard for Cardiac Monitors, Heart-Rate Meters, and Alarms.
1280 x 1024
60 Hz
Manual
Input Voltage
Switching
(VAC)
Required?
160 - 286
n/a
100, 120 -127,
Yes
200, 200 - 240
power dissipation
Specifications
Input
Dissipated
Frequency
Power (max)
(Hz)
(Watts)
50 - 60
50 - 60
Value
Hardware Description
2-51

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3170M3155bM3151bIntellivue seriesM3169aM3154b