Philips M3150B Installation And Service Manual page 422

Information center system
Table of Contents

Advertisement

Troubleshooting
Step 19. Follow the OS installation procedure described in Operating System
HP D4692N Kayak
Step 1. Properly shutdown and restart the PC and IMMEDIATELY press and hold down
XU
Step 2. Select the Advanced menu in the Setup Utility by pressing the keyboard right
Step 3. Select the ISA Resource Exclusion submenu by using the down arrow key and
Step 4. Select IRQ 3 and press Enter to access the IRQ3 submenu.
Step 5. Change the IRQ3 value to Reserved and press Enter - or - if
Step 6. Select IRQ 5 and press Enter to access the IRQ5 submenu.
Step 7. Change the IRQ5 value to Reserved and press Enter.
Step 8. Press the keyboard ESC key once to return to the Advanced menu.
Step 9. Select Integrated I/O Ports and press Enter to access its submenu.
Step 10. Select Serial Port B and press Enter to access its submenu.
Step 11. Change its value to Disabled and press Enter - or - if a 4 Channel Recorder or
Step 12. Select Base I/O address (Serial Port B) and press Enter to access its submenu.
Step 13. Select Interrupt (Serial Port B) and press Enter to access its submenu.
Step 14. Change its value to IRQ3 and press Enter.
Step 15. Press the keyboard ESC key twice to access the BOOT menu.
Step 16. Select Quickboot Mode and press Enter to access its submenu.
Step 17. Change its value to Disabled and press Enter.
Step 18. Press the keyboard ESC key once to access the Exit Menu.
Step 19. Select Exit Saving Changes and press Enter to access the Setup Confirmation
Step 20. Select Yes to Save configuration changes and exit now? and press Enter to
If only changing the BIOS settings, remove the floppy from the PC and reboot. If using the
FIST floppy as part of an installation, the system will then return to the blue HP Kayak PC
Workstation screen and the installation process will continue. When the MS DOS 6.22
Startup Menu reappears:
7-136
Maintenance, Troubleshooting, and Repair
Installation.
the F2 key.
arrow key.
press Enter.
Channel Recorder or external modem is installed change IRQ3 value to Available
and press Enter.
external modem is installed then change its value to Enabled and press Enter.
Change its value to 2E8 and press Enter if no external modem
Change its value to 2F8 and press Enter if external modem present
window.
save the changes.
4

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3170M3155bM3151bIntellivue seriesM3169aM3154b