Step 2. Power-On Self-Test (POST)
The POST checks each of the subsystems.
Warning
If any of the POSTS fail, an Error Message will be displayed and the POST will be
terminated. If an Error Message is displayed, record the error message(s) and contact the
Response Center.
DO NOT USE WITH PATIENTS!
Note
No action is required during this phase.
Step 3. Services Load:
Step 4. Windows Subsystem Start (audible sound)
Information Center Boot Process
Step 5. Initialize Services (grey screen)
Step 6. Starting Services (grey screen)
Listen to the external speaker for the 4 sound tones.
Step 7. Monitoring Mode
Step 8. Turn on the Printer (if included)
The Philips system is now ready for patient monitoring and the following Tests shall be
performed.
Performance
After the Philips system enters Monitoring Mode, Philips patient monitoring software shall be
Test
fully tested as follows:
Step 1. Verify that the date and time are displayed correctly.
Step 2. Verify mouse or trackball functionality as follows.
The OS Services are then loaded.
The Windows subsystem is then initiated and the Microsoft Windows logo appears.
Philips Services are first initialized and a grey screen appears with Waiting for
services to initialize. displayed in the center followed by Entering monitoring
mode....
Information Center Services software are initialized and a grey screen appears with
Testing sound system.... During this test, four distinct tone sequences will be played.
Following the loading of the SDProcess application, the Information Center system will
immediately enter Monitoring Mode and the Main Screen will appear on the Philips
display.
Configuring the System
Start-up and Configuration
8-3