Troubleshooting
Source
Description (Event Code)
SDN
Error: 905, Overflow of one of the driver to card recorder
queues (i.e., too much data sent to a recorder). Data not sent to
the recorder.
Error: 906, Underflow of one of driver to card recorder queues
(i.e., not enough data sent to the recorder).
Error: 907, Unexpected state of the software for one of the
recorders. The software logic for the recorder will be
reinitialized and the driver to card recorder connection will be
reestablished. Some recording data may be lost
Error: 909, Unexpected state of the software for one of the
recorders, The software logic for the recorder will be
reinitialized and the driver to card recorder connection will be
reestablished. Some recording data may be lost.
Error: 910, Unexpected state of the software for one of the
recorders, The software logic for the recorder will be
reinitialized and the driver to card recorder connection will be
reestablished. Some recording data may be lost.
Error: 911, All SDN and recorder message packets were
transferred from the driver to the card correctly but there was
no end of message header. Transfer did not complete.
Error: 912, Unexpected state of the software for one of the
recorders. The software logic for the recorder will be
reinitialized and the driver to card recorder connection will be
reestablished. Some recording data may be lost.
Error: 913, Have reached the end of the driver to card transfer
count while reading an sdn message packet and there was no
end of message header. Transfer did not complete.
Error: 914, Have reached the end of the driver to card transfer
count while reading a recorder message packet and there was
no end of message header. Transfer did not complete.
Error: 917, The card did not have enough time to transfer the
data to the driver across the PCI bus (i.e., some device may be
holding onto the PCI bus for too long). Transfer did not
complete.
Error: 918, Card to driver message queue overflow. The card
was not able to push data to the driver fast enough. One 32ms
frame of data lost.
Error: 919, The driver interrupted the card indicating that the
next card to driver transfer could be started but the transfer
count indicates that the driver did not finish reading data from
the last transfer from the card. Data from the card to the driver
may be lost.
Error: 920, Unexpected state in the card to driver data transfer
software.
Error: 928, Device id list is too small
7-38
Maintenance, Troubleshooting, and Repair
Table 7-8. System Events
Corrective Action
Reboot Philips system. If problem persists,
replace SDN Interface Card
No action required. The Philips system will
automatically correct the situation.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card