Providing Electrical Power

Providing Electrical Power
The following IntelliVue Information Center components are provided with 3-wire,
grounded power cables.
Each 3-wire, power cable must be separately connected to the appropriate 3-wire, grounded
electrical receptacle. See Figures 5-13 and 5-14 for power cables and instrument connection
points
Caution
Warning
IntelliVue Information Center system components do not comply with IEC 60601-1 and
may not be located within the Patient Environment. See Safety in Chapter 4.
5-32
Hardware Installation
Display(s)
Processing Unit
UPS
Power Supply for Philips 2-Channel Recorder Rack
Power Supply for Philips 4-Channel Recorder
HP LaserJet Printer
6-Way Video Splitter
BEFORE connecting power cables to electrical outlets:
1. Verify that all components of the Information Center system that can be
turned off are turned OFF!
2. Review Safety in Chapter 4 to assure proper electrical grounding and
isolation.
3. Set the input voltage selection switch
workstation to the proper line voltage.
4. Set 120V UPS rear panel switch settings to those shown in Figure 2-27.
5. Set Power Distribution Module input and output voltage switch settings to
their proper values. See Table 2.
PC workstation, 60V 2-Channel Recorder Power Supply, and the optional
external modem or 4-Channel 24V Power Supply connect to the BATTERY
BACKUP outlets on the UPS.
Displays, video splitters, and printer hubs may be connected to ACCESSORY
outlets of the UPS or to a separate non-UPS outlet.
HP LaserJet Printers should not be connected to a UPS.