Configuration
using the buttons below the SCCs field. These buttons are the same as for Clinical Units. If there are
only IntelliVue Patient Monitors or M3/M4 bedside monitors on the Network, this field is not
applicable.
Buttons below the field are for adding, changing, and deleting SCCs names.
Button
Description
New
Adds a SCC name to the list.
Edit
Changes the name of a SCC in the list
Remove
Removes a name from the SCC list.
At least one Clinical Unit and must be entered in the Clinical Units and SCC's fields at all
Note
times. Therefore, a new name must be added before an unwanted name is deleted.
List of networked centrals and hallways lists the Information Centers and Clients on
the Network.
Columns in the List of network centrals and hallways field are:
Field
Description
Device Name
Lists the names of the Information Centers and Clients (with an identifying icon) on the Network of
this Server. Device Names are entered in the Add to network domain window when the device is
added to the network.
Note—On the M3170 Patient Link, only the Patient Link device can be configured.
Model #
Model Number of the Device -- M3155 Information Center or M3151 Information Center Client
Name given to the Device when it was installed. Generally it is the same as its Device Name. Host
Host Name
Names for each device must be unique for the Server to properly identify it
Internet Protocol Address that identifies the location of the Device on the Network. It is read from
IP Address
the Device when it connects to the Server for the first time. If no IP Address appears, then the
Device has not yet connected to the Server.
Location
Optional field for identifying the location of the Device in the hospital. It is entered in the Add new
device to this domain window when the device is added to the network
Unit
Clinical Unit in the Clinical Units list served by the Device.
SCC
(for SDN systems) Name of the Serial Communications Controller in the SCCs list that an
Information Center is connected to.
CS#
(for SDN systems) The Central Station number that the Information Center uses to connect to the
SCC.
6-26
Software Installation & Configuration
• Click New to bring up the Add/Edit SCC window.
• Type in the name of the SCC to be added in the Type new name or label: field. The name
has a limit of 12 characters.
• Click on (highlight) the SCC name to changed to select it.
• Click Edit to bring up an Add/Edit SCC window. (Double clicking on the SCC name opens
this window directly).
• Type in the new name of the SCC in the Type new name or label: field.
• Click OK and the changes appear in the SCC list.
• Click on (highlight) the SCC name to be deleted to select it.
• Click Remove and the name will be deleted from the list. If an Information Center is mapped to
this SCC, it cannot be removed until the association is removed.