Description
Source
(Event Code)
ProductConfig
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Could not write value to Registry because Error = XX
Could not read value from Registry because Error = %d
WinSock connection failed
Could not read which configuration applications to run from
the Registry.
Could not start application because Error = XX
PswdScrDlg
Could not start EMS Main Menu application
... Application could not be started.
Logon User does not have the privilege to access EMS
applications. Need to update the access rights of the
SystemManager User.
WinSock connection failed
System manager password entered
Philips Private password entered
RecMgr
Error: 2504, The default case on a switch was reached.
Error: 2510, No primary wave was found in the pds. Using the
first two waves.
Error: 2511, No data was found in the pds. This could be a
temporary startup condition.
Error: 2512, A pds collector could not be created for the
specified patient.
Error: 2513, A patient could not be found from the
information supplied.
Error: 2514, This unmarshall routine does not support the
version of data that was marshalled.
Table 7-7. Application Events
Explanation
Corrective Action
Reboot system. If this persists, reinstall OS
and Philips software
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall OS
software
Reinstall Philips software
Reboot system.
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall OS
software
Verify proper user access
Verify proper user access
Report to Response Center and reboot
system.
Reboot system. If this persists, reinstall
Philips software.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
Reboot system. If problem persists, reinstall
Philips software. If problem still exists,
contact Response Center.
Maintenance, Troubleshooting, and Repair
Troubleshooting
7-31