When the Save process has completed:
Step 7. Remove the Archive disk from the A: drive and store it in the holder on the side of
Service Log data for the first 200 entries have been stored on the disk. If the file was saved
as a dB Export file, the data can now be imported to another database.

Status Log

The Status Log provides information on the operational status of all devices connected to
the same Network as an Information Center or Client.
Note
If the Information Center or Client is in Local Database Mode, data on other Network
connected devices will not be current and may be incorrect.
There are two versions of the Status Log -- Quick Unit Status and All Data Categories --
and both can be accessed from either the All Controls window or the Service menu.
Quick Unit
The Quick Unit Status version of the Status Log provides summary information to clinical
Status
users for quick identification and reporting of problems with devices in their clinical unit. The
Quick Unit Status window is shown in Figure 7-12.
The following information is displayed
Domain Name is a read-only field that displays the Host Name of the Database Server on
the Network of the device. For local database Information Centers, it is blank.
Status Category has a pull down list for selecting the desired version of the Status Log
If the file cannot be saved for some reason, an error message to Cancel the Save
operation is displayed.
Replace the Archive disk with another disk and repeat the process.
the Server.
Figure 7-12 Quick Unit Status Log Window