Activating The Network/Server System - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Buttons at the bottom of Database Server's Service windows, see for example Figure 6-18,
do the following:
Button
Description
Returns to the Service Password window of Figure 6-18 in a locked out mode
Main Screen
requiring the Service password to be reentered
Enter Config Mode
Opens the WARNING window:.
Clicking Yes causes all Information Centers and Clients on the Network to reboot and go into
Local Mode and the Server stops storing patient data.
The remaining buttons indicate the status of Information Centers and Clients on the
the status indicators and their meanings:
OFF-LINE
Indicates the device is off, in Desktop, start-up, or its connection to the Network is not functioning.
When the Server is being configured, all devices are shown as OFF-LINE
MONITORING
Indicates the device is connected to the Server and in Monitoring Mode
SERVICE
Indicates the device is connected to the Server and in Service mode
LOCAL MON
Indicates the device is in Monitoring Mode; storing a limited set of data locally, not on the Server
LOCAL SVC
Indicates the device is not accessing the Server and is in Service mode
Activating the
When all Information Centers and Clients have been configured, the complete system can be
Network/
activated as follows:
Server
Step 1. Reconnect the Network cables of all Information Centers and Clients on the
System
Step 2. Configure Unit Settings for all Information Centers and Clients on the Network.
Note
If clinical Unit Settings have been configured when the Information Center or Client was in
Local Database Mode, they will be lost when the device reconnects to the Network. Therefore,
Unit Settings must be configured after the Network connection has been made.
Network to cause them to reconnect to the Network.
. Following are
Network
Software Installation & Configuration
Configuration
6-59

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3170M3155bM3151bIntellivue seriesM3169aM3154b

Table of Contents