Philips M3150B Installation And Service Manual page 277

Information center system
Table of Contents

Advertisement

Procedure The procedure for configuring the Bed Config window is as follows:
Step 3. Select the desired Fast Alarm Display option -- Disabled, Red and Yellow, or
Step 4. Select the correct Central Station: # for this device for Information Centers. This
Step 5. Verify that entries in the SCC: and clinical Unit fields of Associations are correct.
Step 6. Verify the entry in the Monitoring: box is correct for the Unit listed above.
Step 7. Select the correct
Note
The number of waves can be 3 for 4 Patient Sectors, 2 for 6 - 12 Patient Sectors, 1 for 16
Patient Sectors, and 2 for 16 Patient Sectors with Dual Resting Display enabled.
Step 8. Verify the desired Patient Sector layout (# of Patients and # Columns) in the
Step 9. Click on Assign to access the Resting Display window.
Step 10. Assign Beds from the Available beds: field to Patient Sectors as follows:
When Beds have been assigned to all Patient Sectors:
Step 11. Click OK to return to the Bed Config window.
Step 12. Click Next> to commit the changes and advance to the next configuration window.
Click on the Bed Label in the Monitor/Overview Beds field to highlight it for
removal.
Click Unassign and that bed is removed.
Red Only.
should have already been set on the Network Configuration page at the Database
Server, or earlier in the configuration on a Information Center without a Database
Server.
# of Patients:
# Columns of Patient Sectors
# of Waves Per Sector
and whether to Show Alarm Limits in the Setup Sector Layout fields whether the
Dual Resting Display option should be enabled.
Layout preview display.
Click on a Bed label in the Available beds list.
Move the cursor to the desired Patient Sector.
Click the mouse button when the Bed label is over the desired Patient Sector. The
Bed label will then appear in that Sector. If a Bed label is replaced, it will reappear in
the Available beds: list.
Software Installation & Configuration
Configuration
6-51

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3170M3155bM3151bIntellivue seriesM3169aM3154b

Table of Contents