Information center system
Table of Contents

Advertisement

Configuration
Purchased
Purchased Options and Support Information displays information that identifies
Options and
hardware, software, and purchased options and provides for entering support information for
Support
users. The Purchased Options and Support Information window is shown in Figure
Information
6-9.
Description Fields in the Purchased Options and Support Information are:
Only the Device Name and Clinical Network IP Address fields require entries. Fields with
Note
an asterisk (*) are optional and all others are read-only.
Field
Server Host Name
This Device
Host Name
Device Name
Monitoring LAN IP Address
Note
If the Server has more than one LAN card installed (e.g. for Web Access), select the IP
Address for the primary Clinical Network connection (i.e. not the 2nd LAN connection).
6-16
Software Installation & Configuration
Figure 6-9 Purchased Options and Support Information Window

Description

Read Only for all devices - associated Database Server hostname
• Networked Information Center: M3155 Information Center Network Database
• Database Server: M3154 Large Database Server/M3169 Small Database Server
• Local Database Information Center: M3150 Information Center Local Database
• Client: M3151 Information Center Client
• Patient Link: M3170 Patient Link
Host name of the Device
Clinical name for the Device. Generally, Device Name is the same as Host Name.
Device Name has a limit of 15 characters
For all devices, must be selected from drop-down list. Each device must have its own
unique IP address.