Electrical Requirements; Network Connection; Network Card And Tcp/Ip Settings - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Electrical
The following table gives electrical requirements for the Information Center installations
Requirements
shown in Figure 4-9. To determine similar specifications for Information Center installations
different from those in Figure 4-9, consult Table 4-5 and/or the documentation for the units
that comprise the system.
Table 4-5. Electrical Requirements for Information Center Installations
Input Voltage
System
A
B
C
Network
The Information Center computer obtains its information from patient monitors connected to
Connection
the Serial Distribution Network (SDN) through the System Communications Controller
(SCC). Therefore, an SDN wall box connection to the SCC should be installed near where the
Information Center computer will be located.
Procedures for installing SDN wall boxes and cabling are described in detail in the CMS Site
Preparation and Installation Manual. Refer to that manual for installing SDN wall box
connections. For available SDN installation material options, see M3199AI Options using
UTP Cable.
Clinical Network cabling, wall boxes, and faceplates are generally the responsibility of a
certified cabling installer. However, the network design should give careful consideration to
the locations of RJ-45 wall boxes, both for the clinical equipment -- Information Centers,
Clients, printers -- and for the active Network components that require them. Distances
between wall boxes and the devices that connect to them will depend on the lengths of RJ-45
and other types of cable.
Network
The following features of the Information Center software require a second Network interface
Card and
card (NIC) to be installed (refer to Chapter 2 Hardware Description):
TCP/IP
Settings
The settings required for this NIC card are for connection to the hospital network and are site-
specific. These settings will be entered into the Control Panel during hardware installation.
Physical network connections including cable runs and wall boxes must also be considered.
Connection to the Hospital LAN requires a 100 Mbit connection.
Input
(VAC)
Frequency (Hz)
110 - 127
UPS: 50/60
All others: 50-60
220 - 240
110 - 127
UPS: 50/60
All others: 50-60
220 - 240
110 - 127
UPS: 50/60
All others: 50-60
220 - 240
12-Lead Export
Holter Export
HL7 Export
Web Access
Patient Data Transfer
Manual
Input Power
Switching
(max) (Watts)
Required?
PC & UPS: yes
858
All others: no
PC & UPS: yes
958
All others: no
PC & UPS: yes
858
All others: no
Site Planning and Preparation
Site Preparation
Notes
Assumes single,
medium display, PC,
Recorder, and UPS.
Assumes dual,
medium displays, PC,
Recorder, and UPS.
Assumes single,
medium display, PC,
Recorder, and UPS
4-23