Locating System Components
M3169 Small
The M3169 Small Database Server should be installed in its intended location. This includes
Database
the Server, display, keyboard, and, mouse. The display, keyboard, and mouse are intended for
Server
a work surface.
Installation
Step 1. Install the Small Database Server components in their intended positions.
Step 2. Install the 650 VA UPS with the proper voltage for the Server as follows:
Voltage Select
Switch
Power Cord Connector
Mouse Connector
Keyboard Connector
5-14
Hardware Installation
– Turn Off the UPS On/Off switches.
– Connect the battery wire of the 650 VA UPS.
– Connect the UPS input power cord to a properly grounded electrical output
– Verify PC Voltage select switch setting
– Connect the PC power cord to a UPS battery backup outlet.
– Verify the UPS Switch settings are accurate, see page 2-25.
– Connect the Philips 9 pin UPS communication cable (gray) to the computer
interface port of the UPS and the Serial A port on the rear of the Server. See
Figure 5-14.
LAN Connector
Philips 9 Pin UPS/Server
Communication
Cable (gray)
Figure 5-14 Connections to the M3169 Small Database Server D510 PC
Hospital LAN
Card (Optional)
-
10BT
100TX
Display Connector
COMPUTER
SITE
INTERFACE PORT
WARNING
4 3 2 1
FAULT
OVERLOAD
PROCTECTOR
UPS OUTPUT
120V 50/60 Hz
BATTERY
BACKUP
ACCESSORY
SURGE
PROTECTION
INPUT
TOTAL
OUTPUT
120V-
CURRENT
50-60 Hz
12A MAX