Philips M3150B Installation And Service Manual page 371

Information center system
Table of Contents

Advertisement

Step 4. Browse to the CDROM drive containing the NLS strings and press OK.
Step 5. In the New Version of NLS Catalog drop list, select the latest NLS version. Press
Begin File Transfer. If errors are encountered, the tool displays the error. See Figure
7-59.
A device that continues to fail can be skipped, and that device must be upgraded using the
Manual process procedure given below.
Step 6. When complete reboot the connected M3155 Information Centers and Clients. They
automatically return to monitoring.
Upgrade NLS Catalog - Manual Process
The NLS text catalog files must be retrieved from the IntelliVue Patient Monitor
Documentation CDROM. Performing this procedure drops all devices into non-Monitoring
Mode.
Step 1. Insert the media containing the NLS text catalog files into the CDROM drive.
Step 2. Open Windows Explorer and browse to the CDROM drive.
Step 3. Select the file folder containing the NLS version (e.g. 6_10) and copy (Ctrl + c or
Edit -> Copy) the folder.
Figure 7-58 Browse for NLS
Figure 7-59 NLS Status
Maintenance, Troubleshooting, and Repair
Troubleshooting
7-85